User-friendly

mgpsThe fully automated Digital Control Panel for Oxygen / Nitrous is in sync with changing Technology with the aim to meet the variegated requirements of clients

  • The Oxygen Control Panel shall be of microprocessor based digital display type. Pressure reduction shall be in two stages. Panel shall be integrated with pressure gauges inside panel on downstream of pressure regulator. Panel shall be fitted with standby line regulator. Line regulators shall have pressure relief mechanism for testing and servicing purpose.
  • The panel shall be fully automatic and shall switch over from Bank in Use to Reserve Bank without fluctuation in delivery line pressure, and without the need of external electrical power. After the switch“over, the Reserve Bank shall become the Bank in Use and the Bank in Use shall become the Reserve Bank. The control panel will be powered by a microprocessor. The unit shall be compact and enclosed in NEMA 1 enclosure.
  • A microprocessor circuit board assembly shall provide a relay output to give indication when or just before the manifold switches from one bank of cylinders to another. The switch over shall be mechanically controlled, not electrically.
  • To avoid excess pressure being supplied to the distribution system, a pneumatically relief valve for the line regulator shall be incorporated. An intermediate pressure relief valve shall be installed between the high-pressure regulators and the line delivery regulators.
  • The control panel incorporates six coloured LEDs “ three for the left bank and three for the right bank “ green for bank in use, amber for bank ready and red for bank empty. Both, the left and right bank pressures, and the main line pressure, should be displayed on the front door of the cabinet by means of LEDs. All pressure transducers, micro switches, and display LEDs shall be pre-wired to an internal microprocessor circuit board.
  • All components inside the control panel like pressure regulators, piping and control switching equipment shall be cleaned for oxygen service and installed inside the cabinet to minimise tampering with the regulators or switch settings.
  • The control panel should be made to provide heavy duty with a flow capacity of over as per requirement of the hospital.
  • Panel shall be compatible for interfacing with the Information Management System. (optional)

User-friendlyFully / semi automatic control panel oxygen / nitrous oxide system Control panel have two first stage regulators, each capable of delivering 120 psig outlet pressure. The first stage regulator connected to the second stage regulator, which will deliver an outlet pressure of 60 psig. The first two regulators, meant for first stage, will be capable of switchover system incorporated from Running to Reserve bank due to differential pressure. The control panel provides for two individual content contact pressure gauges to indicate the cylinder pressure in the two wings of the manifold and common pressure gauges to indicate the delivery / line pressure. Control panel have audio alarm for bank changeover. The control panel has built in signal lamp indications for changeover. Control panel covered with suitable cover indicating the respective services.